| Home | E-Submission | Sitemap | Contact Us |  

The Korean Journal of Pathology 1977;11(4): 359-365.
임상화학검사 Kit에 관한 검색
Evaluation of Chemical Kits Available in Korea
ABSTRACT
In order to improve the quality of laboratories, routine use of commercial kit is recommended. This evaluation study was done to see if commercial kits presently available in Korea will suffice to use in routine laboratory. Domestic kits for cholesterol, amylase, alkaline phosphatase, BUN and albumin and imported kit (DADE) for cholesterol were evaluated. Correlation of results of kits and our laboratory results was good except amylase. The test for precision of kits showed higher coefficient of variation in cholesterol, amylase and BUN than acceptables range, but no significant differences from our laboratory results. The test of accuracy for cholesterol and amylase showed that percentages of error of kits were within the acceptable range although higher than our laboratory result in cholesterol determination. In summary the domestic chemical kits of cholesterol, alkaline phosphatase, BUN and albumin seemed to be suffice to routine laboratory tests, however kits for amylase needed further evaluation because of its high coefficient variation.
TOOLS
PDF Links  PDF Links
Full text via DOI  Full text via DOI
Download Citation  Download Citation
CrossRef TDM  CrossRef TDM
  E-Mail
Share:      
METRICS
1,119
View
10
Download
Related articles
Survey for the Laboratory Methods of Chemical Tests in Korea  1979 September;13(3)
Evaluation of cytopathologic diagnosis of lung carcinoma.  1991 June;2(1)