Clinical Experience and Sensitivity of the AutoPap 300 QC System in Cervicovaginal Cytology.
Skip Navigation
Skip to contents

J Pathol Transl Med : Journal of Pathology and Translational Medicine

OPEN ACCESS
SEARCH
Search

Articles

Page Path
HOME > J Pathol Transl Med > Volume 9(1); 1998 > Article
Original Article Clinical Experience and Sensitivity of the AutoPap 300 QC System in Cervicovaginal Cytology.
Sung Ran Hong, Jong Sook Park, Hoi Sook Jang, Yee Jeong Kim, Hy Sook Kim, Chong Taik Park, In Sou Park
Journal of Pathology and Translational Medicine 1998;9(1):37-44
DOI: https://doi.org/
Departments of Pathology, Samsung Cheil Hospital, College of Medicine, Sung Kyun Kwan University.
prev next
  • 1,751 Views
  • 23 Download
  • 0 Crossref
  • 0 Scopus

OBJECTIVE: False negatives of cervical smears due to screening errors pose a significant and persistent problem. AutoPap 300 QC System, an automated screening device, is designed to rescreen conventionally prepared Pap smears initially screened by cytotechnologists as normal. Clinical experience and sensitivity of the AutoPap 300 QC System were assessed and compared with current 10% random quality control technique. MATERIALS AND METHODS: In clinical practice, a total of 18,592 "within normal limits" or "benign cellular changes" cases classified by The Bethesda System were rescreened by the AutoPap System. In study for sensitivity of The AutoPap System to detect false negatives, a total of 1,680 "within normal limits" or "benign cellular changes" cases were rescreened both manually and by the AutoPap System. The sensitivity of the AutoPap System to these false negatives was assessed at 10% review rate to compare 10% random manual rescreen.
RESULTS
In clinical practice, 38 false negatives were identified by the AutoPap System and we had achieved 0.2% reduction in the false negative rate of screening error. In study for sensitivity, 37 false negatives were identified by manual rescreening, and 23 cases(62.2%) of the abnormal squamous cytology were detected by the AutoPap System at 10% review rate. CONCLUSONS: The AutoPap 300 QC System is a sensitive automated rescreening device that can detect potential false negatives prior to reporting and can reduce false negative rates in the laboratory. The device is confirmed to be about eight times superior to the 10% random rescreen in detecting false negatives.